GCMS/FID - Perkin Elmer Clarus

Chromatograph : Perkin Elmer Clarus 690
Detector = Mass Spectrometer: SQ8 - Electron Impact Quadrupole (EI).
Column: Elite-5MS (capillary, methyl-silicone grafted with 5% PH ME siloxane), 60 mx 0.25 mm x 0.25 µm & Elite-Wax (capillary, stationary phase in polyethylene glycol), 40 mx 0.18 mm x 0.30 µm .
Injector : Autosampler (x108).

Chromatograph : Perkin Elmer Clarus 680
Detector 1 = Mass spectrometer: Perkin Elmer Clarus 600T - Electron Impact Quadrupole (EI).
Column : Elite-5MS (capillary, methyl-silicone grafted with 5% PH ME siloxane), 60 mx 0.25 mm x 0.25 µm & Elite-Wax (capillary, stationary phase in polyethylene glycol), 40 mx 0.18 mm x 0.30 µm .
Detector 2 = FID : Autoignition with PPC electronic regulation.
Injector : Autosampler (x84).

Chromatograph : Perkin Elmer Clarus 580
Mass spectrometer : Perkin Elmer Clarus 560 S - Electron Impact Quadrupole (EI).
Column : Elite-5MS (capillary, methyl-silicone grafted 5% PH ME siloxane), 40 mx 0.18 mm x 0.18 µm.

Chromatographe : Perkin Elmer Clarus 680
Mise en évidence de siloxanes



Application domain

Identification and determination of volatile organic compounds (raw materials, solvents, adjuvants, residual monomers, pollution, etc.).

This device makes it possible to carry out a structural analysis when it is coupled to the pyrojector and an ultra-volatile analysis when it is coupled to a headspace.


Principle

Separation and identification technique reserved for the analysis of relatively volatile and thermally stable, non-ionic compounds with a molar mass of less than 650 g/mol.

The volatilizable organic compounds are separated in the chromatograph by traversing the chromatographic column then are identified by the MS detector (mass spectrometry).

When a molecule is introduced into the mass spectrometer, its ionization produces several types of ions. We then obtain a mass spectrum which consists of the abundances of these ions (on the ordinate), as a function of their mass/charge ratio (on the abscissa).
The area of the chromatographic peaks of each compound is proportional to their content in the sample.


Benefits

Fast & simple,
Allows for precise identification (name, CAS number, probability of identification, etc.),
Allows access to the content (quantification, semi-quantification, relative area ratio),
Very sensitive (LQ from ppb to ppm),
Little sample required (a few µg or µL)


Disadvantages

Detection of organic compounds exclusively (non-detection of ionic elements),
Detection range from 100 to 650 g / mol.

GCMS in action:

Application examples : 

Determination of VOCs in polymeric films by GCMS

Identification and quantification of UV stabilizer in ABS granules by Extraction / GCMS

Identification and quantification of extractables in a HDPE film by Extraction / GCMS

OUR ANALYTICAL EQUIPMENT 

During injection into the GC, all organic compounds will be separated by the chromatographic column and then identified by the MS (mass..
GCMS-FID

GCMS-FID

The sample is directly injected into the pyrolyzer oven heated to 550 ° C using a solid syringe. This instantly causes the fragmentation of..
PYROLYZER

PYROLYZER

During the injection into the LC, all the semi-volatile organic molecules will be separated by the chromatographic column and then detected..
UPLC/UV/TOF-MS

UPLC/UV/TOF-MS

During the injection into the LC module, all the semi-volatile organic molecules will be separated by the chromatographic column and then..
UPLC-MS-MS

UPLC-MS-MS

The HeadSpace / GCMS technique does not require preparation. The sample is placed in a gas tight vial and hermetically sealed using a..
HEADSPACE/GCMS

HEADSPACE/GCMS

This spectral analysis makes it possible to know the nature of the covalent bonds present in the sample. Each of them (for example..
IRTF SPECTROMETER

IRTF SPECTROMETER

A SEM (Scanning Electron Microscope) uses an electron beam to obtain enlarged images of very good resolution. It can also be used to..
SEM/EDX

SEM/EDX

CENTRIFUGE

CENTRIFUGE

Certain molecules form by reaction colorimetric complexes having maximum absorption in UV-Visible. The color of these complexes is read..
UV-VIS SPECTROMETER

UV-VIS SPECTROMETER